MD8 AirScan Features


The Sartorius MD8 AirScan air sampler together with gelatin membrane filter disposables constitute an airborne microorganism sampling system for accurate, reproducible and quantitative detection of airborne microorganisms in clean rooms and isolators, whether conventionally ventilated or under laminar flow.

Operational Advantages

The MD8 AirScan offers distinct operational advantages. When used with the isolator installation kit, the filter head is mounted inside the clean room/ isolator with the sampling device mounted outside the room. This is desirable because it eliminates corrosion problems for the air sampler from the sterilizing agents, there is no particle emission or turbulence, no risk of secondary contamination and the filter head requires minimal space inside the critical test area. The sampling head can be positioned in the vicinity of the critical zone and allows for in-situ sterilization of filter holder and connecting hose. Users now have the possibility of operating the sampler from a distance using either of two remote control configurations: a PC and connecting cable; the accessories required include a PC or laptop (Microsoft Windows 95/98 software or higher) and a printer; or a PLC interface unit. This version requires only separate interface port for the PLC unit.

md8
Accuracy
The MD8 AirScan also has several advantages in air sampling accuracy. Very low bacteria counts can be measured because the high flow rates allow sampling of large volumes in a short time, 1 m3 in less than 8 minutes. This makes the method ideal for validating and monitoring the most critical sterile areas. When monitoring clean rooms and isolator equipment with laminar-flow installations, clean benches and air conditioning equipment, it is absolutely necessary that sampling be conducted under isokinetic conditions. This requires that the air intake rate of the air sampler be equal to the laminar flow rate and the sampling head of the gelatin membrane filter to be positioned perpendicularly to the air stream. If sampling is done at a rate lower than the laminar air flow, too many large particles will be collected on the filter surface due to the effects of inertia. Conversely, if sampling is done at a rate higher than the laminar air flow, too many smaller particles will be collected on the filter surface. In either case, the results will not be representative of the particle distribution actually present in the laminar flow system during sampling. The MD8 AirScan utilizes the gelatin membrane filter method which boasts a 99.9% retention rate for airborne bacteria and is the only method capable of reliably collecting airborne viruses. The system is capable of capturing the lowest microbial concentration possible; 1cfu/m3.
isokinetic_sampling

The Gelatin Membrane Filter Method

The MD8 AirScan utilizes the gelatin membrane filter method for collecting airborne microorganisms. This is the only sampling method that boasts a 99.9% retention rate for airborne microorganisms compared with 40%-60% for other collection methods and a 99.9% retention rate for airborne viruses. Due to the porous nature of the filter, all bacteria-laden particles are captured, regardless of size and the residual moisture content of the filter prevents desiccation of the retained organisms. Filters are pre-sterilized (gamma irradiation), single or triple bagged, H2O2 safe and disposable.

direct_method After sampling, there are two filter transfer methods: direct and indirect. Using the direct method, the filter is placed directly on an agar nutrient plate. The gelatin filter dissolves, leaving the microorganisms in direct contact with the plate. Using the indirect method, after sampling, the gelatin filter is dissolved in a sterile solution. The solution is then membrane-filtered and the filter disk is placed on an agar medium. The indirect method is useful for special evaluations, such as when inhibitors are present in the air being sampled, high colony counts are expected or when microorganisms collected are to be incubated on several different agar types simultaneously. When the virus concentration is to be determined, the indirect method must be used so that retained viruses can be cultivated in the solution and analyzed by virological or molecular biological methods.
isolator_3 isolator_accessories
isolator2 isolator1


Specs

MD8 AirScan
Air throughput 2.0 m3/h to 8.0 m3 /h, in increments of 0.1 m3 /h
Max deviation +/- 5% within the temperature air throughput range of15oC to 35oC
Timer 1-99 minutes, adjustable in steps of 1 minute
Power 700 W
Safety fuses 6.3 Amp. 115 V; 3.15 Amp. 230 V
Noise level Max 62 dB(A) with gelatin membrane filters
Weight Approx. 6.5 kg
Dimensions (L x W x H) in mm 375 x 242 x 228
Voltage 100/115/230-volt version (50/60 Hz)
Gelatin Filter Disposables
Materials Membrane: Gelatin, Filter holder: Cyrolite
Filter Diameter 80mm
Thickness approx. 250 um
Pore size 3.0 um
Retentive Capacity

Bacteria: 99.9995% at the inlet velocity of 0.25 m/s

Viruses: 99.9% for T1 phages (coli phages) at a relative humidity of 50% and inlet velocity of 0.3 m/s

99.94% for T3 phages (coli phages) at a relative humidity of 80% and inlet velocity of 0.3 m/s

Max allowable humidity 85%
Max allowable temperature 30oC
Sterilization Gamma irradiation
Calibration Unit
Measuring range min. 1.6 m3/h max 16 m3/h
Max error of measuring system up to 3.2 m3/h = +/- 3%; 3.2-16 m3/h = +/- 1.5%
Current consumed by the electronic circuitry for measurement 8 mA
Pulse output 5,000 pulses= 1m3
Allowable ambient temperature -10oC…+40oC
Allowable temperature of the media -10oC…+40oC
Allowable humidity (rel.%) 30-80%
Type of protection IP 40 during operation with filter holder, connector hose and PVC
Hours of operation with fully charged battery pack approx. 70hrs
Time needed to completely recharge battery 30hrs
Weight approx. 5.6 kg
Dimensions (L x W x H) in mm 295 x 174 x 246

Accessories

-Adaptor for gelatin membrane filter units:
Allows for the attachment of the gelatin membrane filter disposables. Available in aluminum or stainless steel.

-Gelatin Membrane Filter Disposables:

Individually packaged, pre-sterilized and ready to connect the adaptor, each consists of a gelatin membrane filter and holder. Sterilized by gamma irradiation, available single or triple-bagged. Come in boxes of ten.

md8accessories-mod-drza
-Calibration Unit:
The portable Sartorius 16740 Calibration Unit has been specifically developed for the flow rate calibration of the MD8 series of air samplers. By employing this calibration unit, users of the MD8 air samplers are able to precisely calibrate their instruments themselves. Calibration is done by connecting the two units inline and measuring the actual volume of air passing through the Calibration Unit at the indicated flow rates.
calibration_unit

-Set of connectors (for hose)

One connector for air sampler inlet to flexible hose and one connector for flexible hose to filter holder/adaptor

set_of_connectors
-Carrying case for MD8 AirScan carrying_case
Flexible PVC hose with reinforced ends, available in 2 lengths: 2 m and 5 m flexible_pvc_hose

Accessories needed for attachment to isolator/ clean room

isolator_accessories

-Silicone tubing, sterilizable, 1 m in length (2 are needed OR 1 silicone tube and 1 flexible PVC hose) silicone_tubing
-Support stand for positioning filter head in isolator/ clean room, includes clamp and clamp holder support_stand
-Adaptor (DN 25 hose barb to 1” – 1 ½” Sanitary flange) to connect MD8 AirScan to an isolator via silicone tubing and a filter capsule OR adaptor (DN 30 hose barb to 1” – 1 ½” Sanitary flange) to connect MD8 AirScan to an isolator via flexible PVC hose and a filter capsule (2 are needed) adaptor_for_pvc_hose

-Adaptor for gelatin membrane filter units:

Allows for the attachment of the gelatin membrane filter disposables; to be fitted on the MD8 inlet. Available in aluminum or stainless steel.

md8accessories-mod-drza
-Stainless steel holding clamp for 1” – 1 ½” Sanitary flanges (2 are needed) stainless_steel_holding_clamp
-Sartofluor capsule with PTFE membrane and sanitary flange inlet and outlet, for sterile air filtration inserted between the MD8 AirScan and isolator. sarofluor_capsule